Insulet Initiates Voluntary Medical Device Correction for Certain Omnipod® Pods in the U.S. and Affected International Markets
Key Terms
medical device correction regulatory
cannula medical
diabetic ketoacidosis (DKA) medical
continuous glucose monitoring (CGM) medical
Food and Drug Administration (FDA) regulatory
This action is separate from the voluntary Medical Device Correction issued on March 12, 2026 affecting certain Omnipod 5 Pods in the
Insulet identified that some Pods from specific lots may have a small tear in the tubing (cannula) just above the skin, between the Pod and the point where the cannula enters the body. If this occurs, insulin may leak outside of the Pod instead of being fully delivered into the body as intended, potentially leading to under-delivery of insulin.
Individuals using an affected Pod may notice wetness on their skin or Pod adhesive or detect the smell of insulin. However, in some cases, this issue may be difficult to detect and go unnoticed.
If insulin is not delivered properly, users may experience high blood glucose levels due to under-delivery of insulin. In the most severe cases, prolonged and persistent high blood glucose levels can lead to diabetic ketoacidosis (DKA), a serious medical condition that requires prompt medical treatment and can be life-threatening if not treated.
Approximately 7 million Pods are included within the scope of this action, approximately
This issue does not affect continuous glucose monitoring (CGM) systems or CGM readings.
The issue was identified through the Company’s ongoing product monitoring.
Insulet has identified the cause of this manufacturing issue and implemented corrective actions designed to prevent recurrence. In addition, the Company has further strengthened its in-process monitoring and quality controls designed to detect cannula tears of this nature.
Insulet is communicating proactively with affected customers and providing clear instructions to help them identify affected lots, discontinue use of impacted Pods, and obtain replacement Pods at no cost. The Company has sufficient supply available to replace affected Pods and does not anticipate any disruption to product availability.
The
Important Information for Omnipod Pod Users
Customers should visit Check Pod Lot to confirm whether their Pod lot number is included in this voluntary Medical Device Correction and request replacement Pods at no cost. A full list of affected lots is available on this site.
If a Pod from an affected lot is currently in use, customers should discontinue use and replace it with a Pod from an unaffected lot.
Customers in the
Customers outside the
Forward-Looking Statement:
This press release includes certain forward-looking statements within the meaning of the Private Litigation Securities Reform Act of 1995, as amended. Forward-looking statements relate to future events, including statements concerning the Company’s plans or expectations regarding any voluntary medical device correction and effects of any voluntary medical device correction on the Company’s business, operations, and financial performance or guidance, and involve known and unknown risks, uncertainties and other factors, many of which are beyond the Company’s control, that may cause the actual results, performance or achievements of the Company to be materially different from its current expectations, assumptions, plans, guidance, estimates and projections, including (but not limited to) the financial, operational, and reputational impact and costs of any voluntary medical device correction, future actions by the FDA and other regulatory bodies, the possibility that any voluntary medical device correction could subject the Company to disputes, claims or proceedings that may adversely affect its business and financial operation and other factors detailed from time to time in the Company’s reports filed with the Securities and Exchange Commission, including those discussed under “Risk Factors” in the Company’s Form 10-K for the year ended December 31, 2025. The Company encourages you to consider all of these risks, uncertainties and other factors carefully in evaluating the forward-looking statements contained in this press release. The forward-looking statements made in this press release are made only as of the date of this press release, and the Company undertakes no obligation to update them to reflect subsequent events or circumstances except as required by applicable law.
View source version on businesswire.com: https://www.businesswire.com/news/home/20260526676657/en/
Media Contact:
Cristal Downing
Chief Corporate Affairs Officer
pr@insulet.com
Investor Relations:
Clare Trachtman
Vice President, Investor Relations
ir@insulet.com
Source: Insulet Corporation